REGENERATED CELLULOSE SYRINGE FILTERS


DESCRIPTION

    

High resistance in filtration and sterilisation of aqueous and organic samples in HPLC applications.

  • Incorporate a regenerated cellulose membrane fixed without adhesive to the polypropylene housing.
  • Hydrophilic.
  • Low protein adsorption.
  • High flow rate.
  • High total throughput.
  • Low hold-up volumen.
  • Resistant to almost all solvents and aqueous solutions in pH range 3-12.
  • Filter diameter related to the volume to be filtered:
    • <1 ml – 4 mm
    • <5 ml – 15 mm
    • <100 ml – 25 m
  • Pore size of 0.20 µm and 0.45 µm
  • Sterilization by autoclaving at 121 ºC and ethylene oxide.


top ^


TECHNICAL SPECIFICATIONS

Material filter
Regenerated cellulose
Material house
Polypropylene
Pore
0.2
0.45
Diameter
4 mm
15 mm
25 mm
4 mm
15 m
25 mm
Flow rate (1) ml/min
0.5
10
60
1.5
30
100
Flow rate (2) ml/min
1.5
55
160
3.0
105
325
Flow rate (3) ml/min
3.5
140
230
10
280
430
Bubble point (4) bar
2.0
2.0
2.0
3.4
3.4
3.4
Filter diameter (mm)
4
15
25
4
15
25
Housing diameter (mm)
12
25
33
12
25
33
Filter area (cm²)
0.07
1.7
4.8
0.07
1.7
4.8
Hold up volumen (before/after bubble point)
/0.005
0.15/0.03
0.3/0.1
/0.005
0.15/0.03
0.3/0.1
Adsorption (for 0.2 µm)
<10 ug/cm2 with BSA
Connectors
Female luer lock inlet
Male luer lock outlet
Max. operational pressure (bar)
4.5
4.5
4.5
4.5
4.5
4.5
Busrt pressure (bar)
6
6
6
6
6
6
Max. temperature (ºC)
121
121
121
121
121
121

 

(1)  Flow rate for waterl at Ap = 1 bar ( 100kPa, 14.5 psi)
(2)  Flow rate for methanol at Ap = 1 bar ( 100kPa, 14.5 psi)
(3) Flow rate for hexane at Ap = 1 bar ( 100kPa, 14.5 psi)
(4)  Wetting fluid: Isoporpyl alcohol


top ^


APLICATIONS

  • Filtration and clarification of small volumes.
  • Fast and efficient cleaning.
  • Filtration and clarification aqueous, organic solution and mixed solutions.
  • Sterilisation and clarification of cell and protein solutions and biological fluids.
  • HPLC: Filtering aqueous and organic solutions
  • GC: Preparation of samples.

top ^


ORDERING INFO

NON-STERILE

units/ pore
0.20 µm
0.45 µm
4 mm 50
JCR020450
JCR045450
15 mm 100
JCR02015100
JCR04515100
  500
JCR02015500
JCR04515500
25 mm 100
JCR02025100
JCR04525100
  500
JCR02025500
JCR04525500


top ^