GLASS MICROFIBRE SYRINGE PRE-FILTERS
DESCRIPTION
![]() |
Recommended for fast pre-filtration of samples with high particles load. |
- Incorporated glass microfibre filter fixed without adhesive to a MBS polymerisate housing.
- Ready to use syringe filter.
- Hydrophilic.
- High flow rates.
- High total throughput.
- Low hold-up volume.
- Up to 500 ml of sample.
- Large filtration area and an optimized geometry of the filter support.
- Luer Lock outlet making it possible to connect additional syringue fiilters.
- Not autoclavable, sterilization only feasible by gamma irradiation or ethylene oxide.
TECHNICAL SPECIFICATIONS
| Material filter | Glass microfibre Syringe |
|---|---|
| Material house | Meta acrylate butadiene styrene polymerisate |
| Pore | 0.20 |
| Colour code | white |
| Flow rate (1) ml/min | 450 |
| Bubble point | 0.5 |
| Filter diameter (mm) | 28 |
| Housing diameter (mm) | 33 |
| Filter area (cm²) | 6.2 |
| Hold up volumen (before/after bubble point) | 0.5/0.25 |
| Adsorption (2) | 0.8-3 ug/cm2 with BSA, 8-12 ug/cm2 with gamma globulin |
| Connectors | Female luer lock inlet Male luer lock outlet |
| Cytotoxicity | No inhibition with MRC-5 or L-929 cells |
| Endotoxins | < 0.06 EU/ml |
| Max. operational pressure (bar) | 4.5 |
| Busrt pressure (bar) | 6 |
| Max. temperature (ºC) | 50 |
(1)Flow rate for water at Ap = 1 bar ( 100kPa, 14.5 psi)
APPLICATIONS
- Pre-filtration of small volume liquids to avoid saturation of the small-porosity membranes.
ORDERING INFO
NON-STERILE
| Units/Box | |
|---|---|
| 100 | JFV25100 |
| 500 | JFV25500 |